● Measures both AC and DC voltages with oscilloscopes.
● Are typically used to measure high voltages in TV sets, power
supplies and instruments in laboratories.
● Lightweight and rugged.
● 2 available contact tips: round needle tip for normal use and a
special flat spring-type for CRT anode.
| |
|
|
| Input impedance |
500 MΩ |
| Attenuation ratio |
1000 : 1 |
| Accuracy |
DCV (0 ~ 20kV) ± 1%, DCV (20kV ~ 30kV) ± 2% ACV (0 ~ 20kV) typically ± 5% at 50/60 Hz |
| Max. working voltage |
30kV DC or 20kV AC |
| Operating temperature |
0°C ~ 40°C |
| Storage temperature |
-10°C ~ 50°C |
| Cable length |
1 m |
| Weight |
Approx. 260g |
| Safety standards |